AI in medical devices · Expert · 13 min lesson

Vigilance, update release and communication for learning and static models

MDR Articles 2(65), 86 to 89 and Annex I section 23.4, EU AI Act Articles 43(4), 73 and 14, FDA PCCP guidance, 21 CFR 803 and 806

What you will learn

Key takeaways

Free practice question

Norvik is told that a patient whose wrist fracture was missed by the software needed surgery after displacement. The hospital cannot say which model version read the image, and Norvik has overwritten old versions. Norvik first learns of the event on a Monday.

What is the correct handling under the MDR?

Wait for a retraining study to prove that the model caused the miss.

Not quite. MDR Article 87 starts the clock at awareness and does not allow waiting for proof of the cause.

Treat it as a complaint and report it in the next PSUR.

Not quite. A possible serious deterioration meets Article 2(65) of the MDR and needs an Article 87 report, not a periodic mention.

Assess it as a possible serious incident, report within the Article 87 window, and record the missing version identity as a finding.

Correct. Articles 2(65) and 87 of the MDR apply to possible serious deterioration, and ISO 13485 clause 7.5.9 traceability gaps need correction.

Ask the hospital to submit the report.

Not quite. The duty to report belongs to the manufacturer under MDR Article 87.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.