Smart QA Academy  All lessons

AI in medical devices · Expert · 13 min lesson

Building the AI technical file: model card, data sheet and traceability matrix

Regulation (EU) 2024/1689 Articles 10, 11 and 13 and Annex IV items 1 to 4 and 6, ISO 13485:2016 clause 7.3.10, ISO 14971 clauses 7.2 and 7.3, GMLP principles 3, 4 and 9

What you will learn

Key takeaways

Free practice question

Norvale's data scientist describes the training data in the file as 'ECG recordings from partner clinics, cleaned'. The notified body asks how many patients, which sites, how labels were made and what was removed.

What does Annex IV item 2(d) expect?

A datasheet describing provenance, scope, selection, labelling procedures and cleaning methods, with counts.

Correct. Annex IV item 2(d) requires datasheets covering provenance, scope, how data was selected, labelling and cleaning.

Only the name of the cloud storage folder.

Not quite. A folder name gives no provenance or labelling detail, which Annex IV item 2(d) expects.

A promise to prepare the data description after certification.

Not quite. Article 11 paragraph 1 requires the documentation before placing on the market.

Nothing, since cleaned data is self-explanatory.

Not quite. Article 10 paragraph 2 requires documented preparation, labelling and cleaning practices, so nothing is self-explanatory.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.