AI in medical devices · Medium · 14 min lesson

AI Act obligations for providers of medical device AI and the interaction with MDR notified bodies

Regulation (EU) 2024/1689 Articles 8 to 17, 43, 47, 72 and 73 with Regulation (EU) 2017/745 and MDCG 2025-6

What you will learn

Key takeaways

Free practice question

Cardiovia wants to retrain Rhythm-Scan every quarter with new ECG data. The change was not described in the initial MDR assessment.

What should the team do about the retraining plan?

Retrain freely; the MDR certificate covers every future model.

Not quite. Article 43(4) can require new assessment for substantial modifications, so unlimited retraining is not covered by the certificate.

Define a pre-determined change plan in the technical documentation and agree it with the notified body.

Correct. Article 43(4) and Annex IV point 2(f) treat documented pre-determined changes as not substantial modifications.

Stop retraining for good.

Not quite. Nothing in the AI Act forbids retraining; it requires planned, documented control under Article 43(4).

Notify the deployer hospitals only.

Not quite. Deployer notice does not replace the provider's conformity assessment duties under Article 43(4).

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.