AI in medical devices · Medium · 13 min lesson

FDA lifecycle guidance for AI software: performance monitoring and change control

FDA draft guidance on AI-enabled device software functions (January 2025), final PCCP guidance (December 2024), ISO 13485 clauses 7.3.9, 8.2 and 8.5, QMSR

What you will learn

Key takeaways

Free practice question

Northgate's monthly report shows camera C2 sensitivity at 0.87 against an action level of 0.90. The analyst notes it in the meeting minutes and moves on, because pooled accuracy is still above target.

What should have happened under the quality system?

A record opened and linked to complaint review, investigation, containment and corrective action as needed.

Correct. A crossed action level is feedback under clause 8.2.1 and a trend under clause 8.4 that must start the corrective action process in clause 8.5.2.

Nothing, because pooled accuracy is above target.

Not quite. Pooled results can hide a failing subgroup, and the monitoring plan in the draft guidance expects subgroup detection. This follows ISO 13485:2016 and the FDA draft guidance.

Wait for the next quarterly review before acting.

Not quite. Delay contradicts the plan to detect performance change in use and leaves patients exposed, which clause 8.5.2 aims to prevent.

Change the action level to 0.85.

Not quite. Moving a threshold after a breach hides the signal and breaks the control intended by the monitoring plan in the draft guidance. This follows ISO 13485:2016 and the FDA draft guidance.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.