AI in medical devices · Medium · 13 min lesson

AI risk management with ISO 14971 and ISO/IEC 23894: hazards unique to AI

ISO 14971:2019 clauses 4 to 10, ISO/IEC 23894:2023, AAMI TIR34971:2023 and EU AI Act Article 9

What you will learn

Key takeaways

Free practice question

Corvale's risk file lists hazards for PulmoAssist: crash, wrong patient record and slow response. The model was trained on images from fixed scanners only, and clinics now ask to connect portable units.

What should the team do first?

Add a data and distribution-shift hazard, with sequence and harm, to the clause 5.4 analysis and review the intended use.

Correct. ISO 14971 clause 5.4 requires all hazards to be identified, and unfamiliar inputs are a hazard source for a model.

Leave the file unchanged because the software has no defects.

Not quite. Clause 5.4 covers any source of harm, so hazards without a software defect still belong in the analysis.

Rely on the overall accuracy figure as proof of safety.

Not quite. Accuracy is a performance result and does not replace identifying hazards under ISO 14971 clause 5.4.

Ask clinics to avoid portable units in a verbal briefing.

Not quite. A verbal request is not an identified hazard or a verified control under ISO 14971 clauses 5 and 7.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.