AI in medical devices · Medium · 13 min lesson

Data governance: sourcing, representativeness, labelling and data quality for AI devices

EU AI Act Article 10 (Regulation (EU) 2024/1689), GMLP principles 3 to 5, ISO/IEC 5259-2:2024, ISO 13485:2016 clauses 4.2.5 and 7.3

What you will learn

Key takeaways

Free practice question

Nerelo's register entry DS-021 for an ECG archive reads: 'Source: partner hospitals. Size: 22,000 recordings. Labels: expert.' No agreement number, collection period, device or labelling method is given.

What is the main weakness of this entry?

The origin, purpose, permission and preparation of the data cannot be shown, so the dataset cannot support evidence.

Correct. Article 10(2)(b) and (c) expect collection origin and preparation steps to be known and recorded for the dataset.

The dataset is too small for any device.

Not quite. Size is not the issue in the entry; Article 10(2)(b) concerns recorded origin and permission, not a fixed number.

The word expert should be capitalised.

Not quite. Style is irrelevant; the gap against Article 10(2)(c) is the missing labelling method and records.

Partner hospitals should not be used as sources.

Not quite. Hospitals are a normal source; the problem is that Article 10(2)(b) origin evidence is missing.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.