AI in medical devices · Medium · 13 min lesson

AI development lifecycle: requirements, model design and IEC 62304 alignment

IEC 62304:2006+A1:2015 clauses 4.3, 5 and 8, MDR Annex I section 17.2, EU AI Act Articles 9 and 11, GMLP principles 4 and 6

What you will learn

Key takeaways

Free practice question

Halden's plan for the Rhythma arrhythmia model says: 'IEC 62304 does not mention machine learning, so it does not apply to us; we follow our data science practices.' The product is a Class IIa device in the EU.

What is the best assessment of this statement?

Correct, because a standard that does not name machine learning cannot be applied.

Not quite. The standard is a process framework, and MDR Annex I section 17.2 expects a lifecycle, risk management and verification regardless of technique.

Incorrect: the standard is process based, so the activities are applied and adapted to learned components, with the mapping documented.

Correct. IEC 62304 clauses 5 to 9 describe processes that fit learned software, and MDR Annex I section 17.2 expects a documented lifecycle.

Correct, but only for neural networks and not for simpler models.

Not quite. The process framework does not distinguish model types; clause 5 applies to software whatever its internal technique.

Incorrect, because the standard requires a specific neural network architecture.

Not quite. IEC 62304 prescribes processes and records and never prescribes an architecture or algorithm for a software item.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.