AI in medical devices · Medium · 13 min lesson

Verification and validation of AI: performance metrics, test sets and acceptance criteria

ISO 13485:2016 clauses 7.3.6, 7.3.7 and 7.3.10, Regulation (EU) 2024/1689 Articles 9, 10 and 15, GMLP principles 3 to 8

What you will learn

Key takeaways

Free practice question

Corvane reports that its sleep-apnoea screening model reached 97 percent accuracy. The test set holds 970 normal recordings and 30 recordings with the condition. The model flags almost nothing. A reviewer asks whether the claim supports the intended use of finding patients with the condition.

What is the main weakness of the accuracy figure?

With only 3 percent real cases, a model that flags nothing scores high accuracy while missing nearly every patient.

Correct. Accuracy blends the classes, so GMLP principle 6 clinically meaningful goals require sensitivity and specificity for the intended use.

Accuracy is always wrong for classification.

Not quite. Accuracy is valid arithmetic but misleading under imbalance, which GMLP principle 6 tells teams to avoid relying on alone.

Ninety seven percent is too low to be useful.

Not quite. The number is not the issue; the metric cannot show whether patients with the condition are found, per Article 9 paragraph 8 metrics.

The set is too large to be representative.

Not quite. Size is not the defect here; the imbalance hides missed cases, which clause 7.3.6 verification must expose.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.