AI in medical devices · Medium · 13 min lesson

Human oversight and transparency: instructions for use, user interfaces and training

Regulation (EU) 2024/1689 Articles 13, 14 and 26, MDR Annex I section 23, ISO 13485:2016 clauses 6.2, 7.3.3 and 7.3.7, IEC 62366-1, GMLP principles 7 and 9

What you will learn

Key takeaways

Free practice question

Corvane's first Pneumo-Flag screen shows a large green banner over the radiograph with the word normal. Pilot readers accept it without opening the plain image. The design team says the interface is final because it passed software verification.

What is the main weakness against Article 14?

Software verification does not show that the interface supports effective oversight; the banner invites over-reliance and needs a usability evaluation.

Correct. Article 14 paragraphs 1 and 4 point b require interfaces that support oversight and awareness of automation bias.

There is no weakness, because verification covers everything.

Not quite. Verification under clause 7.3.6 checks requirements, not whether users can oversee the system effectively as Article 14 expects.

The banner colour must be red instead.

Not quite. Colour alone does not address automation bias; Article 14 paragraph 4 point b concerns over-reliance in general.

Only the deployer is responsible for the interface.

Not quite. Article 14 paragraph 1 places the design duty on the provider, while Article 26 adds duties for the deployer.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.