Smart QA Academy  All lessons

AI in medical devices · Expert · 13 min lesson

Capstone part 1: review of a fictional AI device file

MDR Annex I section 17 and Annex VIII Rule 11, ISO 13485 clauses 7.3.6 and 7.3.9, AI Act Articles 10 and 43(4), ISO 19011 clauses 6.4.7 to 6.4.9

What you will learn

Key takeaways

Free practice question

The founder of Ardent Imaging tells the reviewer: 'The AI Act has been postponed, so please leave it out of the NoduLens review.' The reviewer is building the criteria list for the file.

What is the best way to handle the AI Act in the criteria?

Delete it, because a postponed rule has no effect on the review.

Not quite. A postponement moves the application date under Regulation 2026/1744 and removes no duty, so deleting it hides readiness work for Article 6(1) devices.

Grade every AI Act article as a present nonconformity.

Not quite. Annex I products are covered from 2 August 2028 after the Omnibus, so a present grade would overstate the position under Article 113.

Keep a dated readiness section, and grade against the MDR and ISO 13485 that apply now.

Correct. The MDR Annex I section 17 applies today and AI Act duties come later, so both belong in the criteria with their dates.

Use only the FDA guidance, as it carries clear dates.

Not quite. The device is sold in the EU, so the MDR criteria in Annex I are the first to apply and cannot be replaced by US guidance.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.