Capstone part 2: findings, risk-based conclusions and an improvement plan

ISO 19011 clauses 6.4.8 and 6.4.9, ISO 13485 clauses 7.3.9 and 8.5.2, MDR Annex I section 17, AI Act timeline under Regulation (EU) 2026/1744

What you will learn

Key takeaways

Free practice question

A draft finding in the NoduLens report reads: 'F-06. The change process is not good. Improve it.' The reviewer has the change record CCR-077 and the procedure number in front of her.

What is the best way to repair this finding?

Add the word seriously before not good.

Not quite. An adverb adds no verifiable fact, while ISO 19011 clause 6.4.8 builds findings on evidence checked against criteria.

Cite clause 7.3.9, name CCR-077 line 8 and state that no assessment preceded release, then grade it.

Correct. The requirement, evidence, statement and grade are the four parts that make a finding answerable under ISO 13485 clause 7.3.9.

Delete the finding, as it is too vague to use.

Not quite. The gap is real and evidenced, so deleting it would breach the evidence based approach in ISO 19011 clause 4.

Add a list of consultants who could redesign the process.

Not quite. Naming consultants is advice and does not state the requirement that is not met under ISO 13485 clause 7.3.9.