ISO 13485 · Expert · 15 min lesson

Auditing the ISO 14971 risk management file

ISO 14971:2019 clauses 4 to 10, ISO/TR 24971:2020, ISO 13485:2016 clauses 4.1.2, 7.1 and 8.2, MDR Annex I section 3

What you will learn

Key takeaways

Free practice question

Marlowe's plan RMP-01 says 'applies during design and development'. The pump has been on sale for two years.

What is the finding?

No finding, because design is the key phase.

Not quite. Clause 4.4 requires a plan that covers life cycle phases beyond design, including clause 10 activities.

The plan does not cover all life cycle phases or the production and post-production activities that clause 4.4 requires.

Correct. A marketed device needs a plan that continues after design. This reading follows the clause structure of ISO 14971:2019 and the guidance in ISO/TR 24971:2020.

A finding against the design lead only.

Not quite. The finding is on the plan content under clause 4.4, not on a person.

No finding, because the plan is approved.

Not quite. Approval does not fix missing content required by clause 4.4. This reading follows the clause structure of ISO 14971:2019 and the guidance in ISO/TR 24971:2020.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.