ISO 13485 · Expert · 14 min lesson

Auditing clean rooms and environmental monitoring

ISO 13485:2016 clauses 6.4.1, 6.4.2, 7.5.2 and 7.6; ISO 14644-1 and -2 (2015); ISO 14644-3 and -5; ISO 19011:2018 clause 6

What you will learn

Key takeaways

Free practice question

Sorrel's class seven clean room certificate states the counts were taken at rest. The room is used for assembly of sterile implant packs, and no operational counts exist.

What is the best audit conclusion?

Compliant, because at rest counts prove the room at all times.

Not quite. ISO 14644-1 separates occupancy states, and clause 6.4.2 needs cleanliness during assembly, not only at rest.

The claim for production is not supported, because the operational state was not tested.

Correct. Clause 6.4.2 requires cleanliness to be maintained during assembly, which ISO 14644-1 operational data would show.

Nonconforming because classification is not required.

Not quite. Clause 6.4.1 requires documented environmental requirements, and ISO 14644-1 classification supports them.

Compliant if the report is signed by the manager.

Not quite. A signature does not provide evidence for the occupancy state required by clause 6.4.2.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.