Smart QA Academy  All lessons

ISO 13485 · Expert · 15 min lesson

Auditing medical device software and QMS software (IEC 62304, cybersecurity, Part 11)

IEC 62304:2006+A1:2015 clauses 5 to 9, ISO 13485:2016 clauses 4.1.6 and 7.5.6, MDR Annex I sections 17.2 and 17.4, FD&C Act section 524B, 21 CFR Part 11

What you will learn

Key takeaways

Free practice question

Tessellate's traceability matrix for release 3.2.0 shows 41 requirements and 38 linked tests. Requirement 27 rounds the insulin dose and is one of the three with no test.

What is the most defensible finding?

Observation only, since coverage is above 90 percent.

Not quite. IEC 62304 clause 5.7 does not accept a coverage percentage when a risk control is unverified.

Nonconformity: a software risk control has no verification before release, against IEC 62304 clauses 5.7 and 7.

Correct. IEC 62304 clauses 5.7 and 7 require risk controls implemented in software to be verified.

No finding, because user acceptance passed.

Not quite. User acceptance is not verification of a specific requirement under IEC 62304 clause 5.7.

A finding against the test tool vendor.

Not quite. The manufacturer is responsible for verification under IEC 62304 clause 5.7, not the tool vendor.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.