ISO 13485 · Expert · 15 min lesson

AI in medical devices I: EU AI Act high-risk obligations and the Digital Omnibus timeline

Regulation (EU) 2024/1689 Articles 6(1), 9 to 15, 11(2), 17, 43 and Annex I; Regulation (EU) 2026/1744; MDCG 2025-6

What you will learn

Key takeaways

Free practice question

Halden's RetinaView is MDR class IIb software and a notified body issued the certificate. The regulatory manager asks whether the AI Act applies.

Which statement can the internal auditor accept as correct?

It is a high-risk AI system because the MDR is listed in Annex I and a notified body is involved, as Article 6(1) requires.

Correct. Article 6(1) joins Annex I coverage and third-party assessment, both met by an MDR class IIb device.

It is out of scope because the MDR already regulates the product completely.

Not quite. The MDR does not exempt the product; Article 6(1) of the AI Act captures Annex I products needing a notified body.

It is high-risk only if the manufacturer registers it in an AI database first.

Not quite. High-risk status under Article 6(1) follows from the legal test and does not depend on registration.

It is high-risk only if more than half of its code is machine learning.

Not quite. Article 6(1) sets no share-of-code rule; the test is Annex I coverage and third-party conformity assessment.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.