ISO 13485 · Medium · 12 min lesson

What a DMR is today - 21 CFR part 820 (QMSR), ISO 13485 clause 4.2.3 and the contents of the device file

21 CFR part 820 (QMSR, in force 2 February 2026) sections 820.7, 820.10, 820.35, 820.45; ISO 13485:2016 clauses 4.2.3, 4.2.5, 7.3.10, 7.5.1

What you will learn

Key takeaways

Free practice question

Brightwell's US quality manager hears that the DMR is gone since the QMSR took effect on 2 February 2026. A colleague proposes deleting the DMR index to save effort.

What is the right response?

Delete it, since part 820 no longer lists a DMR.

Not quite. Deleting removes the specifications that clause 4.2.3 and clause 7.5.1 still require for manufacture.

Keep it only for devices sold in the United States.

Not quite. Clause 4.2.3 applies to every device type the organisation makes, whatever the market where it is sold.

Replace it with the design history file.

Not quite. The design file in clause 7.3.10 does not hold manufacturing procedures and acceptance criteria.

Keep the content, maintained as the clause 4.2.3 medical device file, and rename the procedure if desired.

Correct. FDA removed the term while the elements remain required through clause 4.2.3 of the incorporated ISO 13485:2016.