Smart QA Academy  All lessons

ISO 13485 · Medium · 12 min lesson

Auditing the device file - sampling, traces, common findings, and DMR versus DHR, DHF and technical file

ISO 13485:2016 clauses 4.2.3, 4.2.4, 4.2.5, 7.3.10, 7.5.1; 21 CFR part 820 sections 820.10, 820.35, 820.45; Regulation (EU) 2017/745 Annex II

What you will learn

Key takeaways

Free practice question

An auditor has one hour at Brightwell and ten days of documents to choose from for the Aria device file.

How should she choose the sample?

Take the first ten documents in the index.

Not quite. An arbitrary selection ignores risk and change, which is where clause 4.2.4 control fails most often.

By risk, recent change and past findings, with the rationale recorded in the audit plan.

Correct. ISO 13485 sets no audit sample size in clause 4.2.4 or elsewhere, so a risk-based, justified selection is the defensible approach.

Ask the auditee to choose the best ones.

Not quite. Letting the auditee select documents invites bias and defeats the purpose of independent verification of clause 4.2.4 control.

Only audit documents without changes.

Not quite. Unchanged documents carry the least risk under clause 4.2.4 compared with recently changed ones, where control usually breaks.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.