EU MDR · Medium · 13 min lesson

Clinical evaluation, UDI, EUDAMED and the person responsible for regulatory compliance (Article 15)

Regulation (EU) 2017/745 Articles 15, 27 to 33, 61 and 86, Annexes VI and XIV, Regulation (EU) 2024/1860

What you will learn

Key takeaways

Free practice question

The Arden plate clinical evaluation report says only: 'A literature review was carried out and the plate is safe.' The report lists no search strings, databases, dates or inclusion criteria. The notified body asks how the review can be repeated.

What is the main gap?

The report is too short for a Class IIb device.

Not quite. Length is not the test; Annex XIV Part A requires a methodologically sound and repeatable process.

The search and appraisal method is not documented, so the evaluation cannot be repeated or judged.

Correct. Annex XIV Part A requires a defined, methodologically sound procedure with identified and appraised data.

The report should be written by the distributor.

Not quite. Article 61 places the clinical evaluation duty on the manufacturer, not the distributor.

The report should omit the literature entirely.

Not quite. Annex XIV Part A expects published literature to be identified and appraised as clinical data.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.