Smart QA Academy  All lessons

ISO 13485 · Medium · 13 min lesson

Risk control, residual risk, the risk management file and production and post-production feedback

ISO 14971:2019 clauses 4.5 and 7 to 10, ISO 13485:2016 clauses 8.2.1 and 8.2.2, EU MDR Annex I sections 2 and 4, Article 83

What you will learn

Key takeaways

Free practice question

For the Pulsar P2 free flow risk, Halden considers three controls: a sentence in the manual, an audible alarm when the latch opens, and a mechanical lock that prevents the latch opening while the driver moves. All three are technically feasible.

Which should be the main control under clause 7.1?

The mechanical lock, because inherently safe design ranks first.

Correct. Clause 7.1 ranks inherently safe design first and information for safety last, as MDR Annex I section 4 also requires.

The sentence in the manual, because it costs nothing.

Not quite. Cost does not set the order in clause 7.1, and information for safety is the weakest option.

The alarm only, because alarms are always enough.

Not quite. An alarm is a protective measure and ranks below inherent safety when the stronger option is feasible under clause 7.1.

Any of the three, because the standard sets no order.

Not quite. Clause 7.1 sets a priority order, and MDR Annex I section 4 repeats it.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.