ISO 13485 · Medium · 12 min lesson

Validating a pouch sealer end to end, and auditing clause 7.5.6

ISO 13485:2016 clauses 7.5.6, 7.5.7, 4.1.6 and 7.6; GHTF SG3 N99-10 sections 3 to 6

What you will learn

Key takeaways

Free practice question

IQ-022 report, Orion RS-120 (fictional): 'Item 5 Calibration certificate: to follow - Pass. Item 9.2.3 Heater set to 190 C: display 190, reference probe 193.5 (criterion plus or minus 2 C) - Pass. Deviations: none. Conclusion: installation successful.' OQ is planned for Monday.

What should the reviewer do?

Reject the IQ conclusion: raise a deviation, calibrate and re-verify the controller before OQ starts.

Correct. Clause 7.6 requires calibrated equipment before use and GHTF section 5.3 lists calibration as an IQ topic; 3.5 C is outside the stated criterion.

Approve the IQ and let OQ start; the certificate can be added later.

Not quite. OQ data taken with an uncalibrated reference cannot be defended, and clause 7.6 requires calibration or verification before the equipment is used.

Approve the IQ, but start OQ only after the final report is written.

Not quite. Writing the report first does not fix the open calibration and criterion failure; GHTF section 5.3 needs the IQ items to pass before the next phase.

Change the criterion to plus or minus 4 C so that the item passes.

Not quite. Changing a criterion after seeing the result defeats objective evidence under clause 7.5.6 c and GHTF section 5.2, which want criteria fixed in advance.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.