ISO 13485 · Expert · 13 min lesson

Packaging design and material selection under ISO 11607-1 and sterile barrier systems

ISO 11607-1:2019 and Amendment 1:2023 clauses 4 to 8, ISO 13485:2016 clause 7.5.7, EU MDR Annex I section 11.4

What you will learn

Key takeaways

Free practice question

Veridian's supplier notifies a change of the pouch film coating to a lower cost adhesive. The datasheet values look the same, and the buyer proposes to accept it at once.

What should the quality system require?

Immediate acceptance because datasheet values match.

Not quite. Datasheet equality does not prove seal, ageing or sterilisation behaviour, so clause 7.5.7 is not met.

A documented impact assessment, and testing or revalidation where the change could affect the sterile barrier.

Correct. Clause 7.5.7 of ISO 13485 requires sterile barrier processes to be validated after changes, so an assessment is needed first.

Acceptance if the price falls by at least five percent.

Not quite. Price is not a criterion in ISO 11607-1 clause 5 or ISO 13485 clause 7.5.7.

Refusal of all supplier changes permanently.

Not quite. Changes are allowed; they must be assessed and controlled under ISO 13485 clause 7.4 and 7.5.7.