IEC 62304:2006+A1:2015 clauses 5.5 to 5.8, 6, 7.4, 8.2 and 9; MDR Annex I sections 17.2 and 17.4; MDCG 2019-16 Rev.1; IEC 81001-5-1:2021

What you will learn

Key takeaways

Free practice question

Halden's release record for DoseGuide 4.3.1 states 'No known anomalies'. The problem tracker holds PR-317, open, high severity: the displayed dose is rounded differently from the calculated dose on one phone model. The release manager says open items are tracked separately.

What does clause 5.8 require?

Document PR-317 as a known residual anomaly and evaluate its risk, or fix it before release.

Correct. Clauses 5.8.2 and 5.8.3 require known anomalies to be documented and evaluated for unacceptable risk before release.

Nothing, because separate tracking of open items is acceptable.

Not quite. Clause 5.8.2 requires the release record itself to document the known anomalies, so separate tracking is not enough.

Remove PR-317 from the tracker so the statement becomes true.

Not quite. Deleting a record to match a statement is falsification, and clause 9.5 requires problem records to be maintained.

Release and mention the anomaly only if a customer asks.

Not quite. Clause 5.8.3 requires the evaluation before release, so disclosure on request comes too late.