Smart QA Academy  All lessons

EU MDR · Medium · 12 min lesson

Structure of the MDR technical file - Annex II, Annex III and how notified bodies expect it laid out

Regulation (EU) 2017/745 Article 10(4), Annex II, Annex III, Article 27; Team-NB best practice guidance on technical documentation

What you will learn

Key takeaways

Free practice question

Corvane's Halcyon index lists a PMS report for a class IIa device. The notified body asks where the periodic safety update report is. The quality manager says class IIa files need only a simple PMS report.

What does the MDR require for class IIa?

A PSUR under Article 86, because class IIa devices need a periodic safety update report, not only a PMS report.

Correct. Article 86 applies to classes IIa, IIb and III, while the Article 85 PMS report is for class I devices.

A PMS report under Article 85 is enough for class IIa devices.

Not quite. Article 85 covers class I devices; class IIa falls under the periodic safety update report of Article 86.

No post-market document, as class IIa is low risk.

Not quite. Annex III requires post-market documentation for all non-custom devices, and Article 84 requires a plan.

Only a complaints log kept by sales.

Not quite. A log alone does not meet the plan and report content required by Article 84 and Annex III.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.