MDSAP · Medium · 14 min lesson

Chapter 4 Adverse events and advisory notices - reportability, timeframes by country and recalls

MDSAP Audit Approach Chapter 4 (2 tasks); ISO 13485:2016 clauses 7.2.3, 8.2.2, 8.2.3 and 8.3.3

What you will learn

Key takeaways

Free practice question

A hospital phones Quillon on Monday 3 September: a monitor froze for forty seconds and a nurse delayed a treatment. The investigation closes on 8 October. The device is sold in the United States.

From which date does the United States thirty day period normally run?

3 September, when Quillon became aware of the information

Correct. 21 CFR 803.50 starts the clock at awareness, which MDSAP Chapter 4 task 1 checks against the complaint date.

8 October, when the investigation closed

Not quite. The investigation end does not start the clock under 21 CFR 803.50, and MDSAP task 1 treats waiting as a failure.

1 January of the following year

Not quite. A calendar year start has no role in 21 CFR 803.50 or in the MDSAP Chapter 4 timeframe check.

The date the device was manufactured

Not quite. Manufacture date is not the trigger; awareness of the event is, under 21 CFR 803.50 in MDSAP task 1.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.