Canada - Medical Devices Regulations SOR/98-282, licences, mandatory problem reporting and recalls

SOR/98-282 sections 32, 34, 46.1, 59 to 61.1 and 63 to 65.1, Health Canada MDSAP notice, MDSAP Audit Approach chapters 1, 2 and 4

What you will learn

Key takeaways

Free practice question

Aldercrest adds a third moulding site. Quality says the MDSAP audit is due in eight months, so the licence can stay as it is until then.

What is the right action?

Open a Canada action before the site produces for Canada, check certificate scope and any licence amendment, and block shipments until aligned.

Correct. Section 32 ties the licence to the MDSAP certificate, and the Audit Approach checks that listed sites match those in use.

Wait for the next audit and ship in the meantime.

Not quite. Shipping from a site outside certificate scope leaves the licence unsupported under section 32 and MDSAP expectations.

Tell Health Canada after the first shipment.

Not quite. Notifying after shipping is too late because a significant change may need an amendment first under section 34.

Ignore it because moulding is a subcontracted step.

Not quite. Outsourced processes stay within the quality system under ISO 13485 clause 4.1.5 and MDSAP chapter 7.

Key terms

Official sources

Go further

The full lesson has narrated slides, more scenario questions with explained answers and a simulation.

Open the AcademyWatch on YouTube

Illustrative example. Educational only, not legal advice.